Browse views: by Year, by Function, by GLF, by Subfunction, by Conference, by Journal

ISCEV Standard for full-field clinical electroretinography (2008 update)

Marmor, M F, Fulton, A B, Holder, G E, Miyake, Y, Brigell, Mitchell and Bach, M (2009) ISCEV Standard for full-field clinical electroretinography (2008 update). Documenta Ophthalmologica, 118 (1). pp. 69-77. ISSN 1573-2622

Abstract

This document, from the International Society for Clinical Electrophysiology of Vision (ISCEV), presents an updated and revised ISCEV Standard for clinical electroretinography (ERG). The parameters for flash stimulation and background adaptation have been tightened, and responses renamed to indicate the flash strength (in cd x s x m(-2)). The ISCEV Standard specifies five responses: (1) Dark-adapted 0.01 ERG (rod response); (2) Dark-adapted 3.0 ERG (combined rod-cone response); (3) Dark-adapted 3.0 oscillatory potentials; (4) Light-adapted 3.0 ERG (cone response); (5) Light-adapted 3.0 flicker (30 Hz flicker). An additional Dark-adapted 10.0 ERG or Dark-adapted 30.0 ERG response is recommended.

Item Type: Article
Related URLs:
Additional Information: Author can archive post-print (ie final draft post-refereeing); Publisher's version/PDF cannot be used
Keywords: Clinical standards; Electroretinogram
Related URLs:
Date Deposited: 22 Feb 2010 11:50
Last Modified: 22 Feb 2010 11:50
URI: https://oak.novartis.com/id/eprint/1666

Search